Temperature Scanner Series SC5x
The Temperature Scanner Series SC5x – data logger is all new and revised scanner / data logger. They incorporate the state of the art 4×20 characters backlit LCD display with white over blue characters to make it visible from the distance.
These data loggers measure up to 24 channels of any process parameters or standard RTD and Thermocouple inputs. Scan rate of 1 second per channel gives an economical solution for process monitoring where fast data rate is not required. With internal memory for storing data, it is very useful in continuous monitoring process where data recording is mandatory. It can also connect to the dot matrix printer for online printing of the data.
These data loggers along with AqWireTMseries of Software makes a complete solution for 21CFR Part 11 compliance. User defined Password protected menus does not allow any unauthorized changes either from PC or from key-board. Every changes are reported either online or in offline mode.
It has an option of communication like RS232C, RS485, USB or TCP/IP. It also supports MODBUS ASCII protocol. When used over TCP/IP, it can easily form apart of the large networked data loggers. It also supports G-Tek’s gtekNetTM web based software.
Features
- Up to 24 channels
- Built-in nonvolatile memory of up to 1MB.
- The online printing and storing is possible from 5 seconds up to 18 hours.
- Individual High low alarm setting
- Resolution of 0.1
- 21CFR Part II compliance(Optional)
- USB Interface
* : Features depends on the product ordered. Please refer to the order code (Model No) of your product for exact features and capabilities.
Specification
- Up to 24 channels
- Built-in nonvolatile memory of up to 1MB.
- The online printing and storing is possible from 5 seconds up to 18 hours.
- Individual High low alarm setting
- Resolution of 0.1
- 21CFR Part II compliance(Optional)
- USB Interface
* : Features depends on the product ordered. Please refer to the order code (Model No) of your product for exact features and capabilities.
Model | SC5x Scanner / Data logger |
Product Code | 5x… |
Display and Operator Panels | |
Display Type | 4×20 Characters backlit LCD display white over blue |
Status Indicator | Store LED, Heart Beat LED, Alarm High/Low Indication |
Panel Keys | Front panel KB consisting of 4 keys for programming and configuration setting |
Analog Input Details | |
No Of Inputs | 8/16/24 Channels |
Theromcouple | Types J, K, R, S, T |
RTD | PT-100; DIN 43760; α=0.00385 |
Volt | 0-1V DC |
mA | 4-20mA; 0-20mA (External Shunt Resistance of 50 Ω 0.1%) |
Sensor Type and Range | Refer to the Table 2 |
Accuracy | |
Accuracy | ±0.5% FSD ±1 digit |
Linearization | Polynomial Based Software Linearization |
Resolution | 0.1 |
CJC Error | ±0.9˚C for 0-50˚C |
Data Scan Rate | 1Channel/1Sec |
Protection | |
Input Impedance RTD/TC /Volt | > 20 MΩ |
Input Impedance mA | 50Ω Shunt External |
CMRR | >100 dB@ 50, 60 Hz at 3 Sample per Second |
NMRR | >50 dB@ 50, 60 Hz at 3 Samples per Second |
Maximum Common Mode Voltage | 5V AC |
Isolation Channel – EARTH | 2.5KV, 1 Minute |
Input Protection | 30V AC/DC Max |
Batch Storage / Data Memory Details | |
Start / Stop | Key Press / Remote Start – Stop |
Store Interval | 5 Seconds to 18 Hrs |
Memory Size | 1 MB / Stop on Full / Rollover |
General | |
Termination | Non Interchangeable, Removable Plugs, Individual For Each Input |
Max No Of Input Cards Per Device | 3 (8 channels each) |
Password Protection | Optional |
Data Storage | Refer to Table 1 |
Environmental | |
Temperature | (Operation)5˚C to 45˚C |
(Limiting) 0˚Cto 50˚C | |
(Storage)-20˚Cto 60˚C | |
Humidity | (Operation)10 to 80 % RH Non Condensing |
(Storage)5 to 90 % RH Non Condensing | |
Altitude | <2000 meter |
Power Requirement | |
Supply Voltage (Mains Operated) | 85-264VAC 47-63Hz |
Power | 15W Max With Maximum Configuration |
Transmitter Power Supply | |
No of outputs available | 1 |
Voltage | 12-15V DC |
Max Load current | 30mA |
Short circuit protection | Yes |
Isolation | NA |
Communication | |
Printer Port (Optional) | Centronics |
Communication | EIA Rs485 2 wire / RS232 |
Protocol | Proprietary |
Baud Rate | 9600/19200 |
Parity | None |
Device Address | Yes, User Settable |
USB | Device Mode/Host Mode |
LED Indication | Yes |
Relay Output Details | |
No of Relays | 1 / 8 Relays |
Max. no. of Relay | 8 |
Max. no. of relay group | 8 (any channel/set-point can be assigned to any group) |
Output Type | Potential Free Contact; 1 Form C |
Alarm Set-point | High/Low individually settable for each channel |
Life Expectancy | Mechanical: 107 operations |
Electrical: 105 operations | |
Relay Contact | 1A 230VAC resistive |
Relay Refresh Rate | 5 Sec |
Isolation Relay-Relay | 2.5KV 1 Minute |
Isolation Relay contact – GND | 1.5KV 1 Minute |
Overall Dimension | |
Dimension L x W x D (mm) | 210 x 192 x 96 |
Panel Cutout (mm) | 184 x 92 + 0.5 |
No. of records that can be stored in the Memory | |
Channels | No. of Records |
8 | 3600 |
16 | 1900 |
24 | 1250 |
Sensor | Range |
RTD (Pt-100) | -100 to 600°C |
0-1V | ± 3000.0 |
4-20mA | ± 3000.0 |
0-20mA | ± 3000.0 |
J type TC | 0 to +400 |
K type TC | 0 to +1200 |
R type TC | 0 to +1700 |
S type TC | 0 to +1700 |
T type TC | -100 to +400 |
1. Panel Mounting Clamps: 2 numbers
Part No. 210034
2. Ethernet Cable
Part No.: 236249
3. USB Cable
Part No.: 210027
4. RS232-C Cable
Part No.: 121011
5. RS485 To 232 Converter
Part No.: 420114
6. RS232 To TCP/IP Converter
Part No.: 200572
Accessories
- Up to 24 channels
- Built-in nonvolatile memory of up to 1MB.
- The online printing and storing is possible from 5 seconds up to 18 hours.
- Individual High low alarm setting
- Resolution of 0.1
- 21CFR Part II compliance(Optional)
- USB Interface
* : Features depends on the product ordered. Please refer to the order code (Model No) of your product for exact features and capabilities.
Model | SC5x Scanner / Data logger |
Product Code | 5x… |
Display and Operator Panels | |
Display Type | 4×20 Characters backlit LCD display white over blue |
Status Indicator | Store LED, Heart Beat LED, Alarm High/Low Indication |
Panel Keys | Front panel KB consisting of 4 keys for programming and configuration setting |
Analog Input Details | |
No Of Inputs | 8/16/24 Channels |
Theromcouple | Types J, K, R, S, T |
RTD | PT-100; DIN 43760; α=0.00385 |
Volt | 0-1V DC |
mA | 4-20mA; 0-20mA (External Shunt Resistance of 50 Ω 0.1%) |
Sensor Type and Range | Refer to the Table 2 |
Accuracy | |
Accuracy | ±0.5% FSD ±1 digit |
Linearization | Polynomial Based Software Linearization |
Resolution | 0.1 |
CJC Error | ±0.9˚C for 0-50˚C |
Data Scan Rate | 1Channel/1Sec |
Protection | |
Input Impedance RTD/TC /Volt | > 20 MΩ |
Input Impedance mA | 50Ω Shunt External |
CMRR | >100 dB@ 50, 60 Hz at 3 Sample per Second |
NMRR | >50 dB@ 50, 60 Hz at 3 Samples per Second |
Maximum Common Mode Voltage | 5V AC |
Isolation Channel – EARTH | 2.5KV, 1 Minute |
Input Protection | 30V AC/DC Max |
Batch Storage / Data Memory Details | |
Start / Stop | Key Press / Remote Start – Stop |
Store Interval | 5 Seconds to 18 Hrs |
Memory Size | 1 MB / Stop on Full / Rollover |
General | |
Termination | Non Interchangeable, Removable Plugs, Individual For Each Input |
Max No Of Input Cards Per Device | 3 (8 channels each) |
Password Protection | Optional |
Data Storage | Refer to Table 1 |
Environmental | |
Temperature | (Operation)5˚C to 45˚C |
(Limiting) 0˚Cto 50˚C | |
(Storage)-20˚Cto 60˚C | |
Humidity | (Operation)10 to 80 % RH Non Condensing |
(Storage)5 to 90 % RH Non Condensing | |
Altitude | <2000 meter |
Power Requirement | |
Supply Voltage (Mains Operated) | 85-264VAC 47-63Hz |
Power | 15W Max With Maximum Configuration |
Transmitter Power Supply | |
No of outputs available | 1 |
Voltage | 12-15V DC |
Max Load current | 30mA |
Short circuit protection | Yes |
Isolation | NA |
Communication | |
Printer Port (Optional) | Centronics |
Communication | EIA Rs485 2 wire / RS232 |
Protocol | Proprietary |
Baud Rate | 9600/19200 |
Parity | None |
Device Address | Yes, User Settable |
USB | Device Mode/Host Mode |
LED Indication | Yes |
Relay Output Details | |
No of Relays | 1 / 8 Relays |
Max. no. of Relay | 8 |
Max. no. of relay group | 8 (any channel/set-point can be assigned to any group) |
Output Type | Potential Free Contact; 1 Form C |
Alarm Set-point | High/Low individually settable for each channel |
Life Expectancy | Mechanical: 107 operations |
Electrical: 105 operations | |
Relay Contact | 1A 230VAC resistive |
Relay Refresh Rate | 5 Sec |
Isolation Relay-Relay | 2.5KV 1 Minute |
Isolation Relay contact – GND | 1.5KV 1 Minute |
Overall Dimension | |
Dimension L x W x D (mm) | 210 x 192 x 96 |
Panel Cutout (mm) | 184 x 92 + 0.5 |
No. of records that can be stored in the Memory | |
Channels | No. of Records |
8 | 3600 |
16 | 1900 |
24 | 1250 |
Sensor | Range |
RTD (Pt-100) | -100 to 600°C |
0-1V | ± 3000.0 |
4-20mA | ± 3000.0 |
0-20mA | ± 3000.0 |
J type TC | 0 to +400 |
K type TC | 0 to +1200 |
R type TC | 0 to +1700 |
S type TC | 0 to +1700 |
T type TC | -100 to +400 |
1. Panel Mounting Clamps: 2 numbers
Part No. 210034
2. Ethernet Cable
Part No.: 236249
3. USB Cable
Part No.: 210027
4. RS232-C Cable
Part No.: 121011
5. RS485 To 232 Converter
Part No.: 420114
6. RS232 To TCP/IP Converter
Part No.: 200572
Please Login first to access download section