Circular Chart Recorder
– CR6” Open Frame

The 6” Open Frame Circular Chart Recorder for OEMs is part of G-Tek’s most popular series, CR2010 Pro, offering utmost reliability and versatility at an affordable price. The pressure pen eliminates concerns about ink depletion or pen replacement. Pressure-sensitive papers are used instead of thermal papers to ensure that the marking is retained for a longer period. These chart plate recorders are widely used in low-temperature equipment, blood banks, incubators, and many more places. It takes direct inputs from RTD Pt-100. Standard chart speeds available are 1 day/ 7 days/ 31 days per revolution. These recorders are microcontroller-based. The crystal-controlled chart drive ensures accurate chart speed. Stepper motors for both pen and chart drive ensure no lag and no backlash. With a cutout of 191 x 163 mm and the smallest depth, it is very convenient to use. They come with a universal power supply as a standard feature. A clamp option is also available that holds panels securely, allowing for easy installation and removal.

Features

  • 6” Chart Width
  • Inkless Chart Recorder
  • Pressure Sensitive Paper for long retention of data
  • Single Analog Input channel
  • Direct Input Standard Pt-100 sensors
  • Universal Power Supply 85-264 V AC, 47-63 Hz
  • Battery Backup of more than 7 day*
  • Available chart speed of 1D, 7D, 31D / Rev
  • Fitting Clamp

* Battery life is calculated assuming process parameter does not vary more than 0.15 Deg/Min and recorder is not turned OFF or ON And the sensor input is not alarming condition. (Battery life depends on number of times recorder on/off and sensor input Variation and alarm condition.)

Specification

Model NoCR-2010  Pro Series Circular Chart Recorder
Product code9994x-xx001
Pens
No. of Pens1
Pen markingContinuous
Pen Response Time< 10 Sec. (Full Scale)
Pen ResolutionStepper Motor Controlled better than 0.1% FSD
OvershootNone
Chart
Chart SpeedIn day mode — 1 day/ 7 days/ 31days per revolution
User settingUser programmable
Chart Calibrated Radius/ Size2.3 ” (calibrated radius approx. 59 mm) / Circular, 5.51″ (140 mm)
Chart RangesRefer Table 1 and Table 2
Display, Operator Panels and Inputs
Display TypeNone
Status IndicatorsALM LED – Red LED blinking indicates Alarm high/low condition

BAT LED – Green LED blinking indicates Normal Operation
Red LED blinking indicates Battery Low status
Yellow LED blinking indicates Error condition
Panel KeysFront panel Keyboard consisting of three keys for configuration and calibration
Number of Analog Input1
Analog Input (Sensor)RTD (PT-100)
Sensor Type and RangeRefer to Table 3
Scan RateMains + Battery: Continuous 1 reading per 2 seconds for 1 Day, 7 Days and 31 Days chart speed
Protection
Input Impedance RTD> 20 MΩ
CMRR> 115 dB @ 50, 60 Hz at 20 sample per second
NMRR> 88 dB @ 50, 60 Hz at 20 sample per second
Maximum Common Mode Voltage5V AC
Isolation Channel-EARTHNA
Input Protection30V AC/DC Max

 

Environmental
Temperature(Operation) +5˚C to +45˚C

(Limiting) 0˚C to +50˚C.

(Storage) -20˚C to +60˚C

Humidity(Operation) 10 to 80 % RH Non-Condensing

(Storage) 5 to 90 % RH Non-Condensing

Altitude<2000 meter
Power Requirement 
Mains Supply + Internal Battery Model

 

Power Supply: 85 to 264 VAC ,47-63 Hz.

Battery: 2×1.5 VDC Ultimate Lithium Non rechargeable battery 3500

mAh

Power Rating3 V, 300 mA (Max) ; 1 W Max with maximum configuration
Fuse TypeNone
Battery Reverse PolarityProtected
Battery backup7 days**
Standards
EMI-EMCIEC 61000-6-2/6-3
SafetyIEC 61010-1
Pollution DegreeII
Installation CategoryII
Vibration2g Peak (10Hz-150Hz)
ShockIEC 61010-1
Physical Characteristics
Overall Dimensions Max

L x W x D (mm)

191 x 163 x 66 mm
WeightApprox. 1 kg (with Sensor and battery)
MountingPanel Mount

** Battery life is calculated assuming process parameter does not vary more than 0.15 Deg/Min and recorder is not turned OFF or ON And the sensor input is not alarming condition. (Battery life depends on number of times recorder on/off and sensor input Variation and alarm condition.)

Table 1- Programmable Chart Ranges

Chart Type

xx

Chart Range***
01+50 to -50
02+50 to -100
03+40 to -10
04-50 to +50
05-10 to +32
06+32 to -10
07+34 to -50
08-10 to +40
09-10 to +15
10-10 to +50
110 to 150
12-6 to -90
130 to 100
140 to 300
15+30 to -40
160 to +50

 

*** First value in the chart range is at the center of the chart.

 

Table 2-  Available Chart(s) and their ranges

21Chart Type

xx

Range***Part Description
410 to + 50 (PS)W6 0 +50 PS
42+30 to -40 (PS)W6 +30 -40 PS
43-40 to +25 (PS)W6 -40 +25 PS
44-115 to +50 (PS)W6 -115 +50 PS
45-200 to 0 (PS)W6 -200  0 PS
460 to +60 (PS)W6 0 +60 PS
Others, Please Specify

 

Sensor TypeValid Input RangeResolutionAccuracyLinearization Error
Pt-100 (RTD)#-200 to 850 °C1 %± 0.5 % FSD± 0.06 °C Max

Table 3-  Sensor Type and Range
 

# α = 0.00385

Accessories

Circular Chart Pack of 30 : 1 (Refer Table 2 for the chart range and part no.)

 

Downloads

    Inquire Now

    TOP