Storage Recorder – Datalogging-CR6

The most popular 6” circular chart recorder of CR2010 series, is also available with an internal memory of 64KB to store data up to 14 days at 5minutes store interval. This recorder is specifically tailored to monitor the blood bank refrigerator temperature data.

This recorder also has an input for Door Open switch. Since door opening and closing is very frequent during the removal of the stored blood, it is very easy to monitor how often and for how much time the door was left open with this recorder.

This recorder uses pressure sensitive chart paper so there is no need to replace the pen. The smudging of the ink is also not there, besides since the paper used is pressure sensitive instead of thermal, the life of the recording is very high. Mains failure and return is also logged.

The user can download, view, analyze and generate the reports for data, events like power failure, door open and close etc by BBRChartTM software  interface via USB. The recorder comes with a built-in battery back up to ensure the proper working in case of power failure. Most other features are similar to the standard CR2010 series 6” recorder.

Features

  • 6” Chart Width
  • Inkless chart recorder
  • LCD Display
  • Real Time Clock
  • Data Storage 64KB of memory
  • Event Logging like Power Fail / Door Open / Power Restore
  • storage of 14 days of data at 5 minute store interval
  • USB port to communicate with PC
  • BBR-Chart Application to download data
  • Generate Reports from BBR-Chart
  • Single Analog Input channel
  • Direct Input standard PT-100 sensor
  • Chart Speed of 7D/Rev
  • Universal Power Supply 85-264 V AC, 47-63 Hz
  • Battery Backup up to 72 hours

Specification

Model NoCR2010 Series; BBR Chart Recorder
Product Code*27×03
Recording System
No. of Pens1 – Suitable for marking on Pressure Sensitive Paper
Pen MarkingContinuous
Pen Response Time<5Sec (Full Scale)
Pen ResolutionStepper Motor Controlled better than 0.1% FSD
OvershootNone
Chart
Chart Speed7D/Rev
Chart Calibrated Radius2.3″ (approx. 59mm)
Chart RangesProgrammable
Display, Operator Panels and Input
Display Type3½ digit LCD
Status IndicatorRTC, Batch Status, Battery Status
Panel KeysFront panel KB consisting of 3 keys for programming and calibration
Analog InputRTD PT-100
Sensor Type and RangeRefer to Table 1
Scan RateContinuous 1 reading per second
Memory
Memory Size64KB
Data StorageUp to 14 days at 5min Store Interval
Alarm RecordsYes (Going out of limit and coming back to limit is considered as one alarm record)
Record for Door Open / CloseYes
Mains Failure RecordYes
Memory TypeCyclical – Most recent data is always available
User Settable Parameters for Recording
Alarm Condition Occurrence Time Delay1 to 99 mins; user settable (through PC)

When the temperature remains out of limit for more than set time then only the alarm condition is assumed.

Real Time Clock (RTC)Settable from Keyboard in DD/MM/YY/HR/MN format
Door Open/CloseInput through Potential Free contact.
Communication
Communication TypeUSB 2.0
PC ApplicationBBR-Chart TM to download data; Configure recorder and generate report.
Protection
Input Impedance> 20 MΩ
CMRR>100 dB@ 50, 60 Hz at 3 Sample per Second
NMRR>50 dB@ 50, 60 Hz at 3 Samples per Second
Maximum Common Mode Voltage5 V AC
Isolation Channel – Earth1.5KV 1 minute
Isolation Channel – ChannelNA
Input Protection30 V AC/DC max
TerminationNon interchangeable, Removable Plugs
Environmental
Temperature(Operation)5˚C to 45˚C
(Limiting) 0˚C to 50˚C
(Storage)-20˚C to 60˚C
Humidity(Operation) 10 to 80 % RH Non-Condensing
(Storage) 5 to 90 % RH Non-Condensing
Altitude<2000 meter

Power Requirement
Supply Voltage (Mains Operated)85-264 V AC 47-63Hz
Battery backupYes
DC Adapter OperatedYes
Power15 W Max with Maximum Configuration
Fuse TypeNone
Battery Backup
Battery 7 Ah 12 V, 1.4 A External Lead Acid Battery
Battery ChargerYes
Battery Reverse PolarityProtected
Minimum Back up>72 Hours
Safety
SafetyIEC 61010-1
EMI-EMCEN 61326-1 Class A
Pollution DegreeII
Installation CategoryIII
Vibration2g Peak (10 Hz-150Hz)
ShockIEC61010-1
IP RatingIP50 (Door and Bezel only)
Overall Dimensions
Dimension (L x W x D) mm232x205x117
Panel Cutout (L x W) mm198×169
Bezel (L x W) mm232×205

*Actual specification may vary depending on the optional features installed.
**Refer to the back panel of recorder for exact rating.

Table 1 Sensor Type, Range, Accuracy and Max Error Specification

Input TypeValid Input RangeResolution ChartAccuracy
PT-100(RTD)-100 to +600˚C2%+/- 0.5% FSD

Table 2 Chart type of Chart Recorder

 

Sr. No.

 

Range**

 

Speed

 

Size

 

Part No.

 

Part Description

210 to +10024H6”304001D60100
220 to +15024H6”304002D60150
230 to +20024H6”304005D60200
240 to +30024H6”304008D60300
25-50 to +5024H6”304016D6-50+50
260 to +160 & -1 to 324H6”304004D60160&-1+3
27+90 to +1404H6”3050014H90140
280 to +1007 Day6”307001W60100
29-50 to +507 Day6”307010W6-50+50
30+50 to -507 Day6”307009W650-50
31-100 to +507 Day6”307005W6-100+50
32+50 to -1007 Day6”307008W650-100
33+40 to -107 Day6”307007W640-10
340 to +507 Day6”307004W6050
350 to +107 Day6”307012W6010
36+20 to -10 (PS)7D6”308009W620-10PS
37+50 to -100 (PS)7D6”308008W650-100PS
39+40 to -10 (PS)7D6”308003W640-10PS
40-50 to + 50 (PS)7D6”210042W6-5050PS
50Other Please Specify

** First value in the chart is at the center of the Chart.
** Chart range low and chart speed are fixed as per the order code.
# PS: Pressure Sensitive Chart

Accessories

 

 

  1. USB cable

USB-cable

 

  1. Charts: Pack of 30
  2. Panel Mounting Clamps:

 

Downloads

    Inquire Now

    TOP