Thermal Calibration
We are a NABLÂ accredited calibration laboratory as per international standard ISO/IEC 17025: 2017. Our laboratory is equipped to do calibration services for temperature & humidity devices.
Temperature and humidity are two very crucial parameters for sensitive products like vaccines, pharmaceuticals, petrochemicals, food processing, etc. It is very crucial that these parameters are measured and maintained during the entire chain of process till the last mile delivery. Calibration of these devices assures confidence in the process and also helps meet the regulatory requirements.
We provide thermal calibration services for temperature & humidity sensors with or without indicators, RTD Sensors, any thermocouple type sensor, temperature gauges, temperature data loggers, thermo hygrometers, temperature & humidity indicating devices, and data loggers.
Features
Key Features:
Calibrate up to
- Temperature Sensors (RTD / Thermocouple): -95 to 1200 °C
- Temperature Data loggers: -30 to 50°C
- Temperature & Humidity Data loggers: 10% to 90% RH & 10 to 50 °C
- Thermo Hygrometer: 10% to 90% RH & 10 to 50 °C
Specification
S.No | Discipline / Group | Measurand or Reference Material / Type of instrument or material to be calibrated or measured / Quantity Measured / Instrument | Calibration or Measurement Method or Procedure | Measurement Range and Additional Parameters where Applicable (Range and Frequency) | Calibration and Measurement Capability (CMC) (±) |
---|---|---|---|---|---|
1 | THERMAL- SPECIFIC HEAT & HUMIDITY | Analog / Digital Thermo Hygrometer, Temperature & Humidity sensor with Indicator / Controller / Transmitter, Temperature & Humidity Data Logger @ 25 °C | Using Standard Temperature & Humidity sensor with Indicator and Humidity & Temperature Generator by Comparison Method | 10 %RH to 90 %RH | 0.71 %RH |
2 | THERMAL- SPECIFIC HEAT & HUMIDITY | Analog / Digital Thermo Hygrometer, Temperature and Humidity sensor with Indicator / Controller / Transmitter, Temperature & Humidity Data Logger @ 50 %RH | Using Standard Temperature & Humidity sensor with Indicator and Humidity & Temperature Generator by Comparison Method | 10 °C to 50 °C | 0.45 °C |
3 | THERMAL- TEMPERATURE | RTD / Thermocouple / Transmitter With or Without Temperature Indicator / Controller / Recorder / Scanner / Data Logger | Using 4 Wire SPRT sensor with Indicator, Multifunction Calibrator and Ultra Cool Drywell Bath by Comparison Method | (-) 95 °C to 140 °C | 0.5 °C |
4 | THERMAL- TEMPERATURE | RTD / Thermocouple / Transmitter With or Without Temperature Indicator / Controller / Recorder / Scanner / Data Logger | Using 4 Wire SPRT sensor with Indicator, Multifunction Calibrator and Dry Block Temperature Bath by Comparison Method | 150 °C to 600 °C | 1.61 °C |
5 | THERMAL- TEMPERATURE | Temperature Data Logger with Inbuilt Sensor | Using 4 Wire SPRT sensor with Indicator and Temperature Chamber by Comparison Method | (-) 30 °C to 50 °C | 1.46 °C |
6 | THERMAL- TEMPERATURE | Thermocouple / Transmitter With or Without Temperature Indicator / Controller / Recorder / Scanner / Data Logger | Using R type Thermocouple with Indicator, Multifunction Calibrator and Dry Block Temperature Bath by Comparison Method | 600 °C to 1200 °C | 2.91 °C |